Appendix A: Guidance and Manufacturer’s Declaration
A-2
Operation Manual SurgiVet
®
V3404 Plus
IMMUNITY
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Electrical fast transient/
burst IEC 61000-4-4
IEC 60601 TEST LEVEL A.C. Mains power voltage should be the typical quality
of a:
Commercial environment.
Hospital environment.
•
•
±0.5 kV to ±2 kV for
power supply lines.
±0.25kV to ±1 kV for
input/output lines.
•
•
COMPLIANCE LEVEL
±0.5 kV to ±2 kV for
power supply lines.
±0.25 kV to ±1 kV for
input/output lines.
•
•
Surge
IEC 61000-4-5
IEC 60601 TEST LEVEL
±1 kV differential
mode
±2 kV common
mode
•
•
COMPLIANCE LEVEL
±1 kV differential
mode
±2 kV common
mode
•
•
Voltage dips, short
interruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
IEC 60601 TEST LEVEL
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle.
<40% U
T
(>60% dip
in U
T
) for 5 cycles.
<70% U
T
(>30% dip
in U
T
) for 25 cycles.
<5% U
T
(>95% dip in
U
T
) for 5 seconds.
•
•
•
•
COMPLIANCE LEVEL
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle.
<40% U
T
(>60% dip
in U
T
) for 5 cycles.
<70% U
T
(>30% dip
in U
T
) for 25 cycles.
<5% U
T
(>95% dip in
U
T
) for 5 seconds.
•
•
•
•
Note: U
T
is the A.C. mains voltage prior to application of the test level.