Chapter 15: Specifications
Operation Manual SurgiVet
®
V3404 Plus
15-1
Chapter 15: Specifications
Parameters Monitored
SpO
2
Pulse Rate
Pulse Strength
Heart Rate (ECG)
Pace Pulses
Respiration Rate
Temperature
Graphics Display
Plethysmogram
ECG
Graphics Trend
Oximetry Bargraph
SpO
2
Range:
50-100% SpO
2
Accuracy:
±2% @ 70-100% SpO
2
±3% @ 50-69% SpO
2
Calibration:
Factory calibrated over 50% to 100% SpO
2
using human blood samples to functional
saturation. Test methods available upon request. No in-service calibration required.
Alarm Ranges:
Low 50-100% (steps = 1), and Off
High 50-100% (steps = 1), and Off
At power up, low %SpO
2
limit will default to 85% if the low %SpO
2
limit was set to less
than 85%.
Display:
3-digit 0.56 inch high LED’s (update rate = 1 Hz)
Averaging:
4/8/16 (Fast / Normal / Slow) pulse beat average (selectable)
Pleth Sweep:
12.5, 25 or 50 mm/sec.
Sensors:
Red: 660nm, 2mw (typical)
Infrared: 905nm, 2-2.4mw (typical)
•
•
•
•
•
•
•
•
•
•
•