Chapter 10: Serial Output
Operation Manual SurgiVet
®
V3404 Plus
10-7
Trend Table Data
Trend data is output in tabular ASCII text format, one table per trend data record, starting at the oldest record.
Trend data is stored every 30 seconds. This interval is not user adjustable. A maximum of 21 hours of trend data
storage is possible. Every time the monitor is turned off, then on again, a new block of trend data is started
in trend memory. Thus, data from days ago might be printed, because trend data is remembered when the
monitor is turned off. For this reason, time stamps are stored and output along with trended parameter data.
Each line of text in the trend tables ends with a carriage return, line feed.
Each new block of many trend data records has the following title information:
Trend data records will be output as follows:
Trend Table
Patient_________________________
Date____________________________
HH:MM:SS BPM SpO2 RR/m Temp
00:00:00 XXX XXX% XXX XXX.XºF
00:00:30 XXX XXX% XXX XXX.XºF
00:01:00 XXX XXX% XXX XXX.XºF
00:01:30 XXX XXX% XXX XXX.XºF
00:02:00 XXX XXX% XXX XXX.XºF