Figure 127.
Setting up DC HiPot test for power pack.
14) When the test
Time Remaining
clock runs down to 00:00, a message dialog will
appear telling you that the motor needs to be discharged. Continue to hold down
the test button (or footswitch) until instructed to release it.
15) When testing completes, release the test button (or footswitch) of the power pack
and return the
Voltage
Output Control
knob to its minimum setting.
16) Click on the
Close
button in the test setup window.
17) If test results should be saved, click on the yellow
Save Results
button in the
Main
window
Tests
tab.
CAUTION
Always allow sufficient time for the test winding to completely discharge
before disconnecting the test leads. The recommended practice is to
discharge the winding for a duration of at least four times the duration of
the DC HiPot test for high-voltage windings.
NOTE
For HiPot operation of the Baker AWA-IV 12kV host analyzer alone, the host
analyzer and the power pack must both be powered up.
126
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
Using power packs with 6/12 kV models