NOTE
It is not necessary to run the output control at a high level to determine if the Baker
AWA-IV is working properly.
If the display still shows no voltage bar, call the service department. Use a meter to confirm
the insulation resistance of the device being tested.
Current bar operation can be tested by shorting test lead 1 to the ground lead. Under this
condition, the voltage bar will not move off the zero line and the current bar should rise very
rapidly and activate the HiPot overcurrent trip warning light (HiPot trip). If the HiPot Trip light
does not light, check for open test leads at either test lead 1 or the ground lead (see “Open
ground check” below). If the problem persists, contact the service department.
HiPot overcurrent trip check
1) Either the HiPot trip lamp does not activate under known shorted conditions, or it
will not go out when test is discontinued.
Call the service department immediately for assistance. Please record information off the
unit and the specific problem prior to calling.
Open ground check
The open ground warning prevents testing.
Answer these questions:
• Has the unit recently been moved to a new location where the outlet might not be
grounded?
• Is the unit being operated in a field where the AC power source is unknown?
• Is the unit being operated on a scope cart that has its own outlet or power source?
• Is the unit being operated using a two-wire extension cord?
• Is the unit being operated on a transformer-isolated circuit?
If you answer “yes” to any of these questions, the unit is probably operational and indicating
that there is open AC line ground connection.
For the first three questions above, use an outlet analyzer to ensure proper wiring
connections to the outlet.
Limited output surge waveform
The display shows a limited output (amplitude) surge waveform. The display rises normally,
but stops at some point. Alternatively, you must continually increase the output control for
successive tests to achieve the same output test amplitude.
Call the service department immediately for assistance on this or any other abnormal
condition noted. Please record basic information from the analyzer and the specific problem
prior to calling.
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
151
Baker AWA-IV troubleshooting
Summary of Contents for Baker AWA-IV 12 HO
Page 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Page 2: ......
Page 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Page 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Page 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Page 213: ......