Figure 93.
Safe to turn on notifications—model dependent.
The following sequence of tests will run automatically:
•
Temperature
—The
Temperature/Resistance Test
setup window will open, ready for
the temperature reading to be entered manually. Readings are taken using an
external device. Enter the temperature reading then click the
Accept
button to
acknowledge the entry and inform the software to proceed with the next test. The
analyzer will automatically proceed with the remainder of the tests.
•
Resistance
—If the readings pass per the requirements of the Test ID, the
Temperature/Resistance Test
setup window will close. If low-voltage leads were
used, a dialog box appears to instruct you to switch to the high-voltage leads.
•
MegOhm
—Starts by ramping up all test leads to operating voltage (500 volts for
this example). This voltage will be held for 60 seconds during which the analyzer
watches for overcurrent trips or insulation resistance values below the minimum
MegOhm setting. If a failure is detected, all testing will stop, and the test leads will
be discharged and grounded. You will be given the choice to repeat the test, stop all
testing, or continue to the next test.
Figure 94.
DC Test setup window—appears when DC Tests are executed.
92
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
Test procedures
Summary of Contents for Baker AWA-IV 12 HO
Page 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Page 2: ......
Page 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Page 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Page 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Page 213: ......