Appendix D — Baker AWA-IV specific winding faults
Software fault messages
Status
Failure type
Failure message
PASS
No failure was detected according to the Test
Model given.
FAIL
USER ABORT
A user abort has been detected.
FAIL
EMERGENCY
SHUTOFF
Emergency Shutoff has been detected
Tested
Tested
No message, cannot determine Pass/Fail
because no test criteria was turned on.
FAIL
CANCEL
User cancelled test.
Resistance failure types
Status
Failure type
Failure message
FAIL
DELTA R
Resistance test result: Fail – DELTA R. Delta R
percent is out of tolerance.
FAIL
OPEN LEADS
Test results: Fail-OPEN LEADS. Open Leads
Detected.
FAIL
TOLERANCE
Test Results: Fail-TOLERANCE Resistance(s)
outside of user defined targeted range.
CAUTION
Reistance out of
range
Resistance values must be greater than 0.500
ohms for HIGH VOLTAGE LEADS and greater
than 0.001 for the RES LEADS. You may stop
Testing, reconfigure Test ID with the appropriate
leads, and repeat the test OR continue with the
remaining tests.
CAUTION
MAX R Range
Exceeded
Test Results: Caution – MAX Resistance Range
Exceeded.
Possible Open lead(s).
FAIL
No Resistance
Solution
Resistance values do not have a solvable
solution.
FAIL
NOISY ADC
Instrument detecting excessive electrical noise.
Resistance measurements are unstable. Check
for noise sources nearby (welders, VFDs, etc.) or
a free wheeling rotor.
Summary of Contents for Baker AWA-IV 12 HO
Page 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Page 2: ......
Page 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Page 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Page 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Page 213: ......