Safety
Information about Personal Safety
0.0
syngo CT 2007E
Emotion Operator Manual
02/2008
0.0
A.2-29
0.0
Special instructions on CARE Vision CT
A.2
CARE Vision CT uses scan parameters and operating conditions
which are unique and may require additional care and radiation
protection measures. In any case the scan time should be kept
as short as possible to avoid unnecessary exposure.
A.2
Radiation exposure to patients
A.2
CARE Vision CT applies continuous exposure at moderate mA
levels. Due to the potentially long scan times and the limited
scan volume, the dose for certain slices may increase to levels
significantly higher than those known from standard CT appli-
cations. The patient’s exposure level is usually estimated by
CTDI-vol (which is designed to give the average dose in the
scanned volume).
A.2
Before starting the scan, the dose rate (CTDIvol in mGy per sec-
ond) is displayed on the monitor. During the CARE Vision CT
scan, the accumulated dose (CTDIvol) reflecting the patient
exposure is displayed on the monitor. The display scale ranges
from zero to 3000mGy. If the table is shifted during the exam,
the accumulated dose will be distributed to different slices and
will be lower than indicated by the display. When a new scan is
loaded, the dose display again starts from zero.
A.2
Summary of Contents for SOMATOM Emotion syngo CT 2007E
Page 4: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 ii ...
Page 8: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 vi ...
Page 240: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Console System Description B 5 24 ...
Page 264: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 System Handling C 2 ...
Page 340: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Function Tests Operation D 2 6 ...
Page 426: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Maintenance E 2 ...
Page 442: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Quality Assurance F 2 ...
Page 526: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Camera Test Quality Assurance F 6 6 ...
Page 536: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Technical Specifications G 2 ...
Page 540: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Release Information J 2 ...
Page 542: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Safety Instructions K 2 ...
Page 544: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Forms L 2 ...
Page 546: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 DIP Report M 2 ...
Page 548: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Location of labels N 2 ...
Page 566: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Glossary O 18 ...
Page 574: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Index P 8 ...