0.0
syngo CT 2007E
Emotion Operator Manual
02/2008
0.0
Constancy Test
Quality Assurance
F.3-26
Performing the table posi-
tion test
F.3
With this quality measurement, you test whether the actual
table position matches the table position displayed.
F.3
The
Table Position
test is performed without triggering radia-
tion. It is started with
Go
in the
Quality Constancy
dialog box.
(→
F.3
i
.3
F
After the previous test has been completed, the
Table Position
test is started automatically.
F.3
✧
Mount a ruler at the mobile part of the patient table so that
the 0 mm mark of the ruler is aligned with the stationary
part of the table.
✧
Place the 135 kg weight on the patient table.
F.3
✧
Start measurement with
Go
.
The horizontal table position is set to 0. The table top moves
300 mm in the direction of the gantry.
F.3
Summary of Contents for SOMATOM Emotion syngo CT 2007E
Page 4: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 ii ...
Page 8: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 vi ...
Page 240: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Console System Description B 5 24 ...
Page 264: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 System Handling C 2 ...
Page 340: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Function Tests Operation D 2 6 ...
Page 426: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Maintenance E 2 ...
Page 442: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Quality Assurance F 2 ...
Page 526: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Camera Test Quality Assurance F 6 6 ...
Page 536: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Technical Specifications G 2 ...
Page 540: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Release Information J 2 ...
Page 542: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Safety Instructions K 2 ...
Page 544: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Forms L 2 ...
Page 546: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 DIP Report M 2 ...
Page 548: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Location of labels N 2 ...
Page 566: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Glossary O 18 ...
Page 574: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Index P 8 ...