![Siemens SITRANS FC430 Operating Instructions Manual Download Page 76](http://html.mh-extra.com/html/siemens/sitrans-fc430/sitrans-fc430_operating-instructions-manual_374486076.webp)
6. Connect wires to terminals, wall mount transmitter.
)
)
Active current output
Passive current output
5
Ca+
6
Cp-
4
C
5
C
7. Tighten cable gland.
Note
For Ex versions active or passive current output is preselected at ordering and cannot be
changed.
Non-Ex versions can be connected as either active or passive.
Note
Load
Signal output: < 500 Ω at 14 to 24 VDC (active), 14 to 30 VDC (passive)
Relay output: 30 VAC/VDC, 100 mA
Passive signal input: 15 to 30 VDC, 2 to 15 mA
Table 5-1
HW polling address
Load [Ω] Voltage (active
Ex) [V]
Voltage (active non
Ex) [V]
24 V DC Voltage supply (passive) [V] Ex and
Non Ex
Measurered
100
3
3
17.7
200
5.9
5.9
19.6
500
11.4
14.9
21.0
1000
14.8
19.2
21.6
2000
17.4
20.1
21.9
5000
19.4
20.4
22.2
10000
20.3
21
22.3
20000
20
20.4
22.4
50000
20.4
20.6
22.5
100000 20.6
20.7
22.6
Connecting
5.2 Connecting FC430
FC430 (From firmware 4.0)
76
Operating Instructions, 06/2017, A5E39789392-AA