HMI menu ID Parameter
Description
Default value
[unit]
Value range
Access
level
3.4.1.2
Pickup S1 amplitude
Measured signal level at pick-up
1. The amplitude on pick-up 1
and pick-up 2 should not differ‐
entiate more than 2 mV.
[mV]
Read
only
3.4.1.3
Pickup S2 amplitude
Measured signal level at pick-up
2. The amplitude on pick-up 1
and pick-up 2 should not differ‐
entiate more than 2 mV.
[mV]
Read
only
3.4.1.4
Max. pickup amplitude
diff
Maximum allowed difference be‐
tween Pickup S1 and Pickup S2.
[V]
Read
only
3.4.1.5
Sensor frequency
The resonant frequency of the
sensor. The frequency depends
on sensor size and liquid density.
[Hz]
Read
only
3.4.1.6
Offset
Zero-point offset compensates
for sensor variation due to proc‐
ess conditions.
[kg/h]
Read
only
3.4.2
Temperature monitoring
3.4.2.1.1
Current value
The measured medium temper‐
ature in the tube.
[°C]
Read
only
3.4.2.2
Transmitter electronics temperature
3.4.2.2.1
Current value
The measured internal tempera‐
ture of the transmitter.
[°C]
Read
only
3.4.2.2.2
Minimum
[°C]
Read
only
3.4.2.2.3
Timestamp at minimum Timestamp when minimum val‐
ue was measured
3.4.2.2.4
Maximum
[°C]
Read
only
3.4.2.2.5
Timestamp at maximum Timestamp when maximum val‐
ue was measured
3.4.2.3
DSL temperature
3.4.2.3.1
Current value
The measured internal tempera‐
ture of the DSL.
[°C]
Read
only
3.4.2.4
Sensor frame temperature
3.4.2.4.1
Current value
The measured temperature of
the sensor frame.
[°C]
Read
only
3.4.3
Inputs and outputs
3.4.3.1
Channel 1 – output with HART
3.4.3.1.1
Loop Current [mA]
Default settings
E.9 Maintenance and diagnostics
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA
493