HMI menu ID Parameter
Description
Default value
[unit]
Value range
Access
level
4.2.3
TAG
Enter a unique Short TAG name
for the device / measurement
point.
Up to 8 characters User
4.2.4
HART device Type
Device Identification: Uniquely
identifies the Field Device when
combined with the Manufacturer
Identification and Device Type.
34
Read
oSLy
4.2.5
HART revision
Device Revision Level defined in
HART protocol.
0
Read
oSLy
4.2.6
Response preambles
Response preambles
5
5 to 20
4.2.7
Dynamic variable mapping
4.2.7.1
SV process value
Select the Secondary HART
process value.
Volume flow
● Mass flow
● Volume flow
● Standard
volume flow
● Density
● Standard
density
● Medium
temperature
● Fraction A
● Fraction B
● Fraction A %
● Fraction B %
● Totalizer 1
● Totalizer 2
● Totalizer 3
User
4.2.7.2
TV process value
Select the Tertiary HART proc‐
ess value.
Density
● Mass flow
● Volume flow
● Standard
volume flow
● Density
● Standard
density
● Medium
temperature
● Fraction A
● Fraction B
● Fraction A %
● Fraction B %
● Totalizer 1
● Totalizer 2
● Totalizer 3
User
Default settings
E.12 Communication
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA
521