Error
Diagnosis
Remedy
Measuring deviations, meas‐
uring errors
Inlet and outlet lines insufficient
Change installation (see Determining a location
(Page 19))
Highly turbulent flow profile
Change inlet line, use flow rectifier, reduce diam‐
eter of tube
For mass measurement: medium density
set incorrectly
Correct density according to current medium
(menu 3.1.4)
Hydraulic calibration values faulty
Check hydraulic calibration values CFH and ZPH.
According to the sensor rating plate and correct if
necessary (menus 6.7.2 and 6.7.3).
Electric calibration faulty
Run self-test (menu 2.4.1);
Check device status (menu 2.1): if "Calibration
failure measuring module" is displayed, send in
the complete electronics for repair
Application-dependent calibration values
faulty
Reset zero point correction (menu 6.6.1) to 0 and/
or reset calibration factor (menu 6.7.1) to 1.0.
Alternatively: Make zero correction (menu 6.6.2)
and manual adaptation of the calibration factor
(menu 6.7.1)
Measured value too low at
dosing
Set attenuation too high
Reduce filter time constant (menu 3.1.7.1)
Fault blanking activated
Deactivate noise suppression (menu 3.1.7.3)
Magnetic field frequency too low
Device may not be suitable for this application
Magnetic field frequency too low
Change excitation to Bipolar (menu 6.7.4)
Measured value too low in
pulsed flow
Temporary exceeding of internal working
range
Increase URV (menu 3.1.2)
Unfavorable hydraulic conditions
Install sensor on the suction side of the corre‐
sponding pump or in front of the feeding valve
Unfavorable hydraulic conditions
Change excitation to Bipolar (menu 6.7.4)
Output current does not cor‐
respond to the current flow
and remains at a constant
value
There is a fault, failure signal 3.6, 22 or 24
mA is output (depending on the setting un‐
der menu 4.1.2)
Determine source of fault under device status
(menu 2.1)
Load too high
Reduce total load including the resistance to <
600 Ω
Current output or power supply defective
Replace complete electronics
Communication via HART
not possible
Minimum load of 230 Ω not available
Increase load to at least 230 Ω and connect HART
Modem/Communicator parallel to this load
Communication via PROFI‐
BUS not possible
Device address set incorrectly (assigned
more than once on the same bus or not
configured in the master)
Correct address under menu 4.1.1 or request and
correct by acyclic communication
Current load on bus too high for connected
feeding device
Try disconnecting other devices from the bus
Adapt the feeding device if necessary
Baud rate between master and coupler/link
not correct
Adapt baud rate to coupler/link
More than two bus terminations exist (incl.
coupler and link)
Reduce to two bus terminations
PROFIBUS interface defective
Check device status (menu 2.1): if "COM module
failure" is signaled, replace the complete electron‐
ics
Troubleshooting
10.2 Application Problems
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
81
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...