Table B-2
Status byte formats
Byte
Bit
No.
7 (MSB)
6
5
4
3
2
1
0 (LSB)
5
Quality
Substatus
Limits
These components may adopt the following values in SITRANS F:
Table B-3
Formats of the quality bits
Bit
7
Bit
6
Profile Designation
Meaning
0
0
bad
Measured value cannot be used
0
1
uncertain
Measured value uncertain
1
0
good (not cascade)
Measured value OK
Table B-4
Formats of the substatus bits
Bit
5
Bit
4
Bit
3
Bit
2
Profile Designation
Meaning
0
0
0
1
configuration error
Parameter error, i.e. upper and lower value for meas‐
uring or output scale identical
0
0
1
1
device failure
RAM or EEPROM defective;
Diagnostic bit "Memory error" is also set and the diag‐
nosis message "RAM error" or "EEPROM error" dis‐
played locally
sensor failure
Measurement was not possible;
Diagnostic bit "Memory error" is also set and the diag‐
nosis message "Measuring path fault" displayed locally
0
1
1
1
out of service
Corresponding function block is in "Out of service"
mode (see parameter "MODE_BLK actual")
0
0
0
0
non specific
Measurement contains too many implausible meas‐
ured values, e.g. due to solids or gas bubbles in the
medium.
Diagnostic bit "Memory error" is also set and the diag‐
nosis message "RAM error" or "EEPROM error" dis‐
played locally
0
0
0
1
last usable value
Failsafe mode: the current measured value has been
replaced with the last good measured value
0
0
1
0
substitute set
Failsafe mode: the current measured value has been
replaced with the agreed failsafe value
0
1
0
0
initial value
Failsafe mode or state before the first measurement:
the current measured value has been replaced with
the start value
0
1
0
0
sensor conversion not accurate
Measured value is outside the sensor limits
(Table 5-1, Page 30)
O.K.
Measured value is OK (normal state)
0
0
0
1
update event
A parameter with the "static" memory attribute has
been changed locally or on the PROFIBUS
Profibus Communication
B.5 Status bytes
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
109
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...