2.5.1
Flow
Simulation of a flow value; affects all out‐
puts, totalizers, limit values and display.
"F" flashes in the top right of the display
during simulation and "Simulation" is dis‐
played under device status.
Value
Simulation value in % of full scale value
0 %
-110% to +110%
Time
Duration of simulation
After expiration of this time or with input of
"End" the normal measuring mode is re‐
sumed
End
• End
• 10 min
• 30 min
• 60 min
2.5.2
Dig. Out.1
Simulation of output signal at the digital
output 1
End
• End
• 0.1 Hz
• 1 Hz
• 10 Hz
• 100 Hz
• 1 kHz,
• 10 kHz
• Alarm on
• Alarm off
2.5.3
Dig. Out.2
1
Simulation of the output signal at the digi‐
tal output 2
End
• Alarm on
• Alarm off
2.5.4
Analog Out.
1, 2
Simulation of the output signal at the ana‐
log output
4 mA
• 0 mA
• 4mA
• 10mA
• 12mA
• 20mA
• Failure signal
1)
Parameters apply only to devices with 20 mA/Hart communication and digital output 2 (7ME5034-0xxxx)
2)
Parameters apply only to devices with 20 mA/Hart communication and digital input (7ME5034-2xxxx)
Parameters
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
95
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...