5.2.1
Product type
Transmitter order number
(7ME5034-
xxxxx-xxA0)
5.2.2
Serial number
Transmitter serial number
(N1xxxx-82xxxx
x)
5.2.3
Software Rev.
Software version in the transmitter
(3.x.x)
5.2.4
Device ID
Clear device identification;
corresponds to HART long address
(Device-specific
HART long ad‐
dress)
1 ... 99 999 (Can only be set
with factory code by HART)
5.2.5
Manuf. Date
Transmitter manufacture date; form:
DDMMYY
(Device-specific
manufacture
date)
DDMMYY (Can only be set
with factory code by HART)
5.2.6
Sensor
Input only possible if no data can be read
from SmartPLUG
5.2.6.1
Nom.Diameter
5
Diameter
Inside nominal diameter of the sensor with
probes: interior diameter of the surround‐
ing tube
(Nominal width)
0.1 m
DN2 ... 2000 / 0.5 ... 80 in,
probe
(Only after ’Probe’ is
selected:) 0,1 ... 5 m
5.2.6.2
Product Type
4
Sensor order number
(7ME5034
-xxxxx-xxA0)
From SmartPLUG
5.2.6.2
Analog TRANS‐
MAG
5
Select ’Yes’ for measuring sensors with an
R100 constant
In this case, a CFH value must be entered in
the 6.7.2 menu according to the following
formula:
CFH = 176,715 * DN2/R100,
DN = nominal diameter in mm)
No
• No
• Yes
5.2.6.3
Serial Number
4
Serial number of sensor
(Serial number)
From SmartPLUG
4)
Parameters apply only to devices with 20 mA/Hart communication and digital output 2 (7ME504x-0xxxx)
5)
Parameters apply only to devices with 20 mA/Hart communication and digital input (7ME504x-2xxxx)
Service
Table A-6
Function group Service parameters
6.1
Enter Code
Input of code number agreed under "Cus‐
tomer Code" (menu 6.2) for releasing local
parameterization
0
0 ... 9999
(according to setting
under menu 6.2)
6.2
Customer Code
Selection of personal code:
• Code 0: Parameters are not protected by
a code.
• Code >0: Parameters can only be
changed after entering the code under
"Enter Code" (menu 6.1)
0
0 ... 9999
6.3
Service Code
Input of factory code for extended service
functions
0
0 ... 99999
6.4
Reset
Device reset (without parameter changes) Cancel
• Cancel
• Reset
Parameters
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
101
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...