3.3.2
Engr. Unit
Engineering unit of volume
or
Engineering unit of mass
When you select a unit of mass, entry of the
actual medium density is mandatory
(menu 3.1.4)
DN 2...12: l
DN >12: m
3
• l, hl, m
3
, Ml, ft
3
, Gal,
MGal, ImpGal, MImpGal
• kg, t, g, lb
3.3.3
Alarm Limit
-1 000 000 l
-108 to +108
3.4
Total. net
Totalizer forw./rev.
3.4.1
Set net
Reset totalizer to zero and stop or start
Cancel 1,2
Totalize 3
• Cancel 1,2,
• Reset+Stop,
• Reset+Start 1,2,
• Totalize 3
3.4.2
Engr. Unit
Engineering unit of volume
or
Engineering unit of mass
When you select a unit of mass, entry of the
actual medium density is mandatory
(menu 3.1.4).
DN 2...12: l
DN >12: m
3
• l, hl, m
3
, Ml, ft
3
, Gal,
MGal, ImpGal, MImpGal
• kg, t, g, lb
3.4.3
Lo Alarm Limit
Lower alarm limit (the lower alarm limit
must be smaller than the upper limit.)
-1 000 000 l -
Frequency 108 ... +108
3.4.4
Hi Alarm Limit
Upper alarm limit (the upper alarm limit
must be greater than the lower limit.)
+1 000 000 l
-108 ... +108
3.4.5
Hysteresis
Hysteresis for limit values
0 l
0 ... +108
(Unit the same as the coun‐
ter unit in the menu 3.4.2)
Device Outputs
Table A-4
Function group Device outputs parameters
4.1
Analog Out
1, 2
Analog output 4 to 20mA with HART com‐
munication
4.1.1
Current limit
Upper limit of analog current
22 mA
20 ... 22.5mA
4.1.2
Error signal
Output current in case of failure.
With the setting "Hold ..s" a current of 3.6
mA is only output in the case of a longer
failure, shorter failures are bridged with
output of the last valid current
3.6 mA
• 3.6mA, 22mA, 24mA,
• Hold continuous,
• Hold 5s, 20s, 40s, 60s,
120s, 240s
4.1.3
Current range
Selecting the current range at the analog
output on which the measuring range
(menu 3.1.2) is shown.
For the setting 0 to 20 mA the failure signal
(menu 4.1.2) must be set to 22 or 24 mA.
4 ... 20 mA
• 4 ... 20 mA;
• 0 ... 20 mA
Parameters
TRANSMAG 2 with MAG 911/E
98
Operating Instructions, 03/2021, A5E00102775-AC
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...