When "Split value active" is selected, the output is active when the current level of the flow value
is within the lower sub-domain 0 to "Split value". In this case the output current is scaled to the
0 to "Split value" range. If the flow value exceeds this sub-domain, the output current is scaled
to the 0 to URV range.
Note
The relay contact is open in the "no function" setting.
• Signal type, menu item 4.3.2
You can set the function of the relay as follows in this menu item:
1. When the relay need to be closed in the event of an alarm:
Flow in forward direction
Limit value message
2. When the relay need to be open in the event of an alarm:
Flow in forward direction
Limit value message
Table 8-3
Relay status
Setting
No alarm
Reverse flow
No limit signalling
Alarm message
Forward flow
Limit signalling
1. Contact closes
2. Contact opens
9
˖
5L
①
Electronic fuse
②
External supply
Figure 8-8
Digital output 2
NOTICE
At high inductive loads, the digital output must be protected against transient overvoltage with
a suitable external recovery diode.
The electronic fuse is tripped in the event of overloading. The recovery time of the fuse is a few
minutes. The relay contact is open in the no-load state.
Functions
8.5 Function Group Device Outputs
TRANSMAG 2 with MAG 911/E
70
Operating Instructions, 03/2021, A5E00102775-AC
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...