10.3
Error messages
Errors are indicated by a flashing "D" (Device error) or "F" (Measurement error) in the top right
hand corner of the local display.
Table 10-2
Error messages
Error
Diagnosis
Remedy
D
Flashing "D" in the top
right hand corner of the
display
General device error (see also section
5.3.2)
Menu 2.1 Call device status
D
Device status (menu 2.1)
"Memory failure"
Internal memory defective
Replace complete electronics
D
Device status (menu 2.1)
"COM Module failure"
HART interface not available in the trans‐
mitter
Replace complete electronics
PROFIBUS interface not available in the
transmitter
D
Device status (menu 2.1)
"Software failure"
General software error
Do a restart (menu 6.4);
If error persists: have firmware reloaded (by serv‐
ice personnel), alternatively: replace complete
electronics
D
Device status (menu 2.1)
"Sensor failure"
Magnetic field current fuse defective
Change magnetic field current fuse (see section
6.3.1)
Magnetic current cable defective (at ter‐
minals 5 and 6 in remote design)
Check cable and change if necessary
Magnet coil defective
Check on remote design:
Switch off device, disconnect magnetic field cur‐
rent cable, measure resistance at terminals 5 and
6 on sensor: setpoint: 8 ... 20 Ω.
Otherwise: change sensor
Reference coil improperly connected (ter‐
minals 55, 66)
Check cable and replace if necessary
Electronics defective
Change complete electronics
D
Device status (menu 2.1)
"Flow measurement dis‐
turbed"
No measurement possible (group mes‐
sage)
Change complete electronics
"SmartPLUG read failure"
Connected sensor without SmartPLUG or
defective SmartPLUG
Check parameterization (range, unit etc.)
A correct measurement is still possible when the
calibration values CFH and ZPH match the values
on the rating plate of the sensor.
Check whether this is the case and correct if nec‐
essary under the menus 6.7.2 and 6.7.3
F
Flashing "F" in the top
right hand corner of the
display
General incorrect measurements, proc‐
ess errors or incorrect parameterization.
Menu 2.1 Call device status
Troubleshooting
10.3 Error messages
TRANSMAG 2 with MAG 911/E
82
Operating Instructions, 03/2021, A5E00102775-AC
Summary of Contents for SITRANS F TRANSMAG 2
Page 6: ...Table of contents TRANSMAG 2 with MAG 911 E 6 Operating Instructions 03 2021 A5E00102775 AC ...
Page 92: ...Dimension drawings TRANSMAG 2 with MAG 911 E 92 Operating Instructions 03 2021 A5E00102775 AC ...
Page 104: ...Parameters TRANSMAG 2 with MAG 911 E 104 Operating Instructions 03 2021 A5E00102775 AC ...
Page 126: ...Index TRANSMAG 2 with MAG 911 E 126 Operating Instructions 03 2021 A5E00102775 AC ...