SIREMOBIL Compact
SPR2-130.061.02
Page 18 of 28
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
6 - 18
Replacing boards / replacing components
•
Configure the following parameters in User setup for MEMOSKOP:
Display of realtime clock
Configure Yes. Select No only upon customer request.
Display of rotated image
Configure Yes.
Monitor zoom
Configure No.
Monitor split
Configure Yes.
Monitor text
Configure Yes.
Maximum image number
Configure 100.
Disable dose rate selection
in user setup
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on disk" and
menu "Dose on Hardcopy output" in User setup are visi-
ble. Otherwise, set to "Yes".
Unit serial number
Program the serial number of the SIREMOBIL Compact.
Do not enter the first zero.
Product name
SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default Program changeable
Default: Yes.
Depending on country specific regulations configure Yes
or No.
Create External disk
Configure No.
Default LUT
Configure 2. Program a different LUT curve only upon
customer request.
Default edge
Configure 1. Program a different Edge enhancement
value only upon customer request.
Software backward compatible Configure No.
Operating mode Parameters
Change the programmed parameters only upon cus-
tomer request.
VCR output
Configure to monitor A.
Hardcopy Release
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Split screen
Configure horizontally. Configure vertically only upon
customer request.
Display of time
Configure Yes.
Set Date Time
Set the current date and time.