Replacing boards / replacing components
6 - 7
Siemens AG
SPR2-130.061.02
Page 7 of 28
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Removing the old I.I.
6
•
Turn the C-arm so that the I.I. is on top.
•
Loosen the I.I. cover and remove it.
•
Rotate the camera to the zero position (6/Fig. 2).
•
Remove the cover adjacent to the I.I. above the high voltage section on the outer side of
the C-arm.
•
Remove the high voltage section and disconnect all cables.
•
Dissipate any residual voltage in the cables by shorting them to ground.
•
Remove the counterbalancing weight on the 7"/17 cm I.I. (1/Fig. 4).
•
Remove the plug from the control section on the I.I. mini voltage supply (1/Fig. 1).
•
Disconnect the protective conductor connections and the grounding screw
(5/Fig. 1 / 1/Fig. 2).
•
Remove plug D4.X1 of the VIDEOMED DC (2/Fig. 2).
•
Secure the end of the cable which you have wound up on the cable reel of the Videomed
DC with adhesive tape.
•
Remove the cable between the VIDEOMED DC and the Compact optics and cut the ca-
ble ties securing it.
•
Remove plug X45 of the cassette contact (2/Fig. 2).
•
Remove both jacks D4.X1 and X45 from the receptacle (2/Fig. 2).
•
Pull the "Anode, E3 and Penning" cables leading from the high voltage section to the I.I.
out of the C-arm (3/Fig. 2).
•
Loosen the retainer clip above the camera and remove it (2/Fig. 1 or 2/Fig. 4).
•
Remove the control panel of the I.I. mini voltage supply from the 9"/23 cm I.I.
•
Loosen the camera set screw slightly and remove the camera (1/Fig. 3).
•
Loosen the Compact optics (3/Fig. 3).
•
Remove the retaining clips on the cables (4/Fig. 1).
•
Remove the O-ring and the cap nut from the anode cable.
•
Disconnect plug M30.X1 (4/Fig. 3).
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
Do not loosen the eccentric screws (Fig. 3/2) on the edge of the
Compact optics! These are used to center the Compact optics to
the I.I. output screen. When removing the Compact optics, make
sure that no dust or dirt contaminates the I.I. output window or
the Compact optics.
WARNING
NOTE