Replacing boards / replacing components
6 - 15
Siemens AG
SPR2-130.061.02
Page 15 of 28
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
MEMOSKOP C-E
•
Replace the MEMOSKOP C-E.
The attachment screws are on the underside of the storage device.
•
Check the software status of the MEMOSKOP. Download the system software status to
the MEMOSKOP if necessary.
•
Use the MEMOSKOP Service Software to import the language file.
•
Configure the MEMOSKOP C-E for the correct video frequency and the number of moni-
tors present.
Refer to Chapter 5 "MEMOSKOP C-E without Keyboard, Emulation ...", "Configuring the
Video frequency" and "MEMOSKOP-Configuration for the number of existing monitors".
•
Configure the following parameters in Technical Setup for the MEMOSKOP:
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP. The MEMOSKOP software status is displayed in the
2nd line of the User Setup. The first number describes the
MEMOSKOP type (2=CE; 3=CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB
& MOD).
Read and record the parameters for MEMOSKOP C-E if you are
able to select User setup. After replacing the MEMOSKOP, con-
figure the MEMOSKOP C-E according to the parameters recorded.
Number of monitors
Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
Hardcopy
Check whether 5000 is programmed in "4 Max Hard-
copy time". If not, program this value.
Edge 1 coef set
Check whether 3 is programmed, if not, program this
value.
Edge 2 coef set
Check whether 8 is programmed, If not, program this
value.
Motion detection setup
Check whether the following are programmed MD1:
Threshold 100 and with MD2: Threshold 120.
If not, program these values.
Date / Time format
Set to country-specific format.
Keyboard
Set to the keyboard layout being used.
Display of realtime clock
Configure Yes. Configure No only upon customer
request.
Display of rotated image
Configure Yes.
Monitor zoom
Configure No.
Default LUT
Configure 2. Program a different LUT curve only upon
customer request.
NOTE
NOTE