
Adjustments / Programs
5 - 23
Siemens AG
SPR2-130.061.02
Page 23 of 30
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
•
Start the service program.
•
Select the Adjustment, Calibrations menu.
•
Click "Yes" to answer the query.
•
In the Combo box "Function Group:," select "Diamentor".
•
In the window "Available Functions", select the line "Start Diamentor Test" and click on
"Execute".
•
Close the status window by clicking on the OK button.
•
Respond the query "Do you want to reset the unit" by clicking "No".
•
Press the -0- key (reset key for fluoroscopic time).
•
The tested Qt*Kg value is displayed on the monitor of the SIREMOBIL Compact.
•
This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: + 5%.
•
By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is dis-
played again on the monitor.
•
The potentiometer CAL on the electronic system of the area dose product measurement
device can be used to adjust the test value to the calculated and programmed Qt*Kg val-
ue.
•
After the adjustment, select "Stop Diamentor Test" on the service PC and click on "Exe-
cute".
•
Close the status window by clicking on the OK button.
•
Respond the query "Do you want to reset the unit" by clicking "No".
•
The test is ended.
•
Check the accuracy of the area dose product measurement device. Refer to the next
section.
Checking the accuracy of the area dose product measurement
device
5
Prerequisites
5
The calibration of the area dose product measurement device has already been per-
formed.
Preparations
5
•
Attach and center the center cross on the I.I. (Fig. 6).
•
Select survey format.
•
Select KV mA curve plateau HC1.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING