Replacing boards / replacing components
6 - 5
Siemens AG
SPR2-130.061.02
Page 5 of 28
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
SIREPHOS
6
•
Open the SIREPHOS cover (see Chapter 1, "Covers").
•
Remove the laser targeting device, if present.
•
Remove the dose measurement chamber, if present.
•
Remove the collimator.
•
Disconnect the SIREPHOS plug.
•
Remove the ground screw for the protective conductor.
•
Remove the rubber stop of the C-arm on the SIREPHOS side. The screws that secure
the SIREPHOS are now visible.
•
Remove the SIREPHOS screws. When doing this, have a second person hold the X-ray
tube unit.
•
Place the new X-ray tube unit on the guide bolts and secure it with both Allen screws.
•
Ensure that all cables are routed properly.
•
Reconnect the SIREPHOS plug and secure it.
•
Reconnect the protective conductor using the ground screw.
•
Do not install the rubber stop yet.
•
Reinstall the collimator and connect it.
•
Reinstall the dose measurement chamber / laser targeting device and connect them.
•
Perform the kV offset adjustment.
•
Perform the generator adjustment (filament circuit learning program).
•
Ensure that the collimator is functioning properly and is set correctly.
•
Ensure that the collimator blades are centered in the blanking circle.
- If possible, slightly loosen the Sirephos screws and tilt the SIREPHOS to adjust it.
- Retighten the SIREPHOS screws.
•
Ensure that the collimator blades of the iris diaphragm are visible during fluoroscopy in
the survey format and zoom format on at least two sides of the monitor image.
•
Check the display of the blades on the monitor screen and adjust if necessary.
•
Check collimation for direct exposure if there is a cassette holder.
•
Reinstall the rubber stop.
•
Ensure that the area/dose product measurement device is functioning properly
(if present).
•
Check the dose rate settings and adjust them if necessary.
•
Check the settings of the maximum skin dose rate.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING