SIREMOBIL Compact
SPR2-130.061.02
Page 30 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 30
Adjustments / Programs
After selecting the desired video frequency, press "Return".
This configures the video frequency you have just selected.
After switching the SIREMOBIL OFF and ON again, the MEMOSKOP will start up at the
correct video frequency.
MEMOSKOP, configuration for the number of existing monitors
5
When you replace the MEMOSKOP, it will be configured for 2 monitors.
This configuration displays USER SETUP and TECHNICAL SETUP on the second moni-
tor.
Therefore, USER SETUP and TECHNICAL SETUP are not visible for SIREMOBIL Com-
pact systems with only one monitor.
Without installed MEMOSKOP keyboard.
•
Use the MEMOSKOP service program to modify the configuration.
With installed MEMOSKOP keyboard proceed as follows:
•
Switch the SIREMOBIL Compact on.
•
On the MEMOSKOP, reconnect the video cable from OUT 1 to OUT 2.
•
Press CTRL + T at the same time on the MEMOSKOP keyboard;
TECHNICAL SETUP is called up and is now visible on the monitor.
•
Call up the menu "Number of monitors" by pressing the "3" key and subsequently the
"RETURN" key on the MEMOSKOP keyboard.
•
With the -> and <- cursor keys, set the number to 1 monitor and press the "RETURN"
key. The text display disappears on the monitor.
•
Press the "RETURN" key once more.
•
Reconnect the monitor to the OUT 1 video output.
The text display is now visible on the monitor.
•
Exit TECHNICAL SETUP by pressing the "HOME" key on the MEMOSKOP keyboard.
MEMOSKOP, configuration and country-specific language (User Setup)
5
Start the program and transfer the country-specific language to SIREMOBIL Compact via
the service PC.
Key "->":
(Start)
"->"
"->"
"->"
"->"
...
Video frequency
50 Hz
60 Hz
100 Hz
120 Hz
50 Hz
etc...
Key "<-":
"<-"
"<-"
"<-"
"<-"
"<-"
"<-"