SIREMOBIL Compact
SPR2-130.061.02
Page 12 of 28
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
6 - 12
Replacing boards / replacing components
•
If necessary, adjust the camera with the clamping ring (1/Fig. 7) by rotating it to the cor-
rect position.
•
Install the VIDEOMED DC in the "0" position on the Compact optics and secure it
(6/Fig. 6).
•
Connect the VIDEOMED DC - Compact optics cable and secure it with cable ties.
Make sure that the cable does not protrude past the edge of the board and that it lies in
the recess of the VIDEOMED DC board.
•
Insert the anode cables for the high voltage panel back into the C-arm (3/Fig. 5).
•
Reinstall the O-ring and the cap nut on the anode cable.
The cable length should be 70 mm from the beginning of the white wire to the cap nut.
•
Reinstall the high voltage panel in the C-arm and reconnect all the cables.
•
Mount the retainer clip above the VIDEOMED DC.
•
Reconnect both jacks, D4.X1 for the VIDEOMED DC and X45 for the cassette contact to
the retainer clip.
•
Connect plugs D4.X1 for the VIDEOMED DC and X45 for the cassette contact.
•
Connect plug M30.X1.
•
Connect plug D120.X1 to the control panel of the I.I. mini voltage supply.
•
Reconnect the protective conductor with the ground screw (5/Fig. 5).
•
Reinstall the cable clamps (3/Fig. 6).
•
Attach 2.1 mm Cu and 60 mm Al prefiltration to the radiation port of the SIREPHOS.
•
Attach the resolution test to the I.I. input screen and check the optical resolution. Read-
just it if necessary.
•
Remove the resolution test from the I.I. input screen.
•
Check the dose rate and readjust it if necessary.
Checking the camera optics centering at the I.I. output
6
•
Mark the center of the I.I. input as precisely as possible using a small washer and adhe-
sive tape. If there is a laser light localizer, the I.I. attachment with the lead cross can also
be used.
•
Select the survey format.
•
Move the camera into the 0° position.
•
Switch fluoro on briefly.
•
Mark the center of the washer or lead cross on the monitor (see Fig. 9).
•
Rotate the camera in the -220° position.
•
Mark the center again on the monitor (see Fig. 9).
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING