4 - 1
Siemens AG
SPR2-130.061.02
Page 1 of 2
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Loading the operating software
4
Control board D1, Host
4
•
Connect the service PC to the serial service interface on SIREMOBIL Compact.
•
Start the service program.
•
Insert the "Parameter" diskette in drive A.
•
Select the Download menu and start the download (ARC / Fil. / Host).
•
After the download has successfully been completed (1 minute), wait for the system to
boot.
•
The SIREMOBIL Compact is operational.
Storage of the error log as ASCII file
4
•
If you store the error log using the backup function of the service program, the error log
will be stored as binary file and can't be read with a word editor.
For easier evaluation of the error log, it is also possible to store it as an ASCII file.
For easier service support from the USC or HSC, please send the error log as an ASCII
file.
Prerequisites:
The service PC is already connected, the service program is already started.
•
Select the menu "Diagnostics...".
•
Select the menu "Error log".
•
Click on the button "Get from unit". Wait until the error log is transferred to the service
PC.
•
Click on the button "Export to file".
•
Select your hard disk drive for storage of the file.
•
For the file name, please enter the current date, the part number and the serial number of
the unit as described in the example.
•
Store the file on the hard disk drive of the service PC to a path of your choice.
•
Exit the service program.
If you cannot run download (e.g. power failure during download):
- Switch the SIREMOBIL Compact off and back on
- Restart download
The parameter diskette must be in drive A before the start of the
service program and during the service work. Without a para-
meter diskette in drive A, the short error messages will not be
stored in the error log file and the text "no error text available" is
stored behind the error codes.
Example:
The current date is (DD-MM-YY):
The unit Part No. is:
The unit Serial No. is:
Enter the file name:
29-03-01
37 76 494
01234
22-03-01_37 76 494_01234.txt
NOTE