5 - 1
Siemens AG
SPR2-130.061.02
Page 1 of 30
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Adjustments / Programs
5
Generator, kV offset adjustment
•
Connect the service PC to the serial interface on board D1.
•
Start the service program.
•
Select the Diagnostic menu.
•
Select the Monitoring menu.
•
Select "KV offset adjustment in the "Group 1" Combo box.
•
Click on "Update Stop".
•
After the "Start fluoro" message in the ACTIONS line, press the radiation release button
on the SIREMOBIL Compact hand switch.
•
Continue to hold the radiation release button as long as the message "Continue fluoro" is
displayed in the ACTIONS line.
•
Release the radiation release button as soon as the message "Stop monitoring" appears
in the ACTIONS line.
•
Click on "Update Stop".
•
Perform the generator adjustment (filament circuit learning program).
The kV offset adjustment has to be followed by the generator
adjustment (filament circuit learning program).
Always perform the kV offset adjustment prior to beginning the
generator adjustment.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
After the "KV error break" message in the STATUS line, click on
"Update Stop". Wait for approx. 10 seconds.
Click on "Update Start" again. After the "Start fluoro" message in
the ACTIONS line, press the radiation release button on the
SIREMOBIL Compact hand switch. Refer to the text which follows
for additional procedures.
NOTE
WARNING
NOTE