6 - 10
Replacing boards / replacing components
Replacing the Compact optics
.
6
Removing the Compact optics
6
•
Turn the C-arm so that the I.I. is at the top.
•
Loosen and remove the I.I. cover.
•
Rotate the camera in the "0" position (6/Fig. 6).
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
3
6
3
2
5
4
5
1
5
1
2
4
4
3
5
3
1
2
3
1
3
2
4
2
5
3
WARNING