SIREMOBIL Compact
SPR2-130.061.02
Page 4 of 28
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
6 - 4
Replacing boards / replacing components
Power supply M13, +12.3 V VIDEOMED DC
6
•
Replace the power supply.
•
Check the power supply voltage and adjust it if necessary. Refer to Chapter 3 of these in-
structions and check the operating voltages.
I.I. mini voltage supply
6
•
Replace the control panel of the I.I. mini voltage supply (installed above Videomed DC).
•
Replace the high voltage panel of the I.I. mini voltage supply.
(Mounted on the C-arm, accessible after removing the cover near the I.I., on the outer
edge of the C-arm. Refer to Chapter 1, "Information regarding this document").
•
Refer to the I.I. test protocol for the voltages E1 / E2 / E3 and A. Check and adjust them,
if necessary, according to Chapter 3, "Checking the operating voltages".
•
Check the functions and adjustments of the collimator. Readjust if necessary.
•
Check the display functions and setting of the blades on the monitor. Adjust if neces-
sary.
•
Check the overall resolution according to the IQ Quick test.
•
Perform the IQ Quick test.
Collimator
6
•
Replace the collimator.
•
Ensure that the collimator is functioning properly and is set correctly.
Adjust it if necessary.
•
Ensure that the collimator blades of the iris diaphragm are visible during fluoroscopy in
survey format and zoom format on at least two sides of the monitor image.
•
Check the collimation for direct exposure if a cassette holder is being used.
•
Check the display of the blades on the monitor and adjust if necessary.
•
Remove the old set of collimator labels near the SIREPHOS and attach the new labels
included in the same position.
Check that the original collimator labels are the same as the new ones.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
WARNING
WARNING