SYSTEM DESCRIPTION
3
65
8026365/V1-0/2022-03|SICK
Subject to change without notice
S U P P L E M E N T A R Y O P E R A T I N G I N S T R U C T I O N S | VMS4200-x/5200-x MID
3.9.5
Test instructions for market surveillance
This section describes the procedure for inspecting the certified VMS5200x- MID
measurement system with regard to:
•
The performance of metrological measurements with defined test objects and the
display of the measurement results on the LFT display.
•
The display of data sets stored in the alibi memory on the LFT display.
•
The display of system component firmware versions via the LFT display.
▸
Establish the voltage supply to the devices via the power supply unit in the cabinet.
•
All system components start up automatically and are then ready for use.
•
The
MEASUREMENT
tab is active in the menu bar. No measured values are available
yet.
▸
Measure test objects made from a dimensionally stable material of various shapes and
surface qualities.
▸
Perform the measurements.
NOTE!
The nominal dimensions (length, width, height) must cover approximately 90%,
50%, and 10% of the total measuring range for each dimension. The actual dimensions
must not be less than the minimum dimensions or exceed the maximum dimensions
specified for the VMS5200-x MID.
▸
Ensure that the required minimum distance of 50 mm is maintained between the test
objects.
▸
Compare the actual values for each measured test object with the dimension values on
the LFT display.
NOTE!
•
Valid measurement results have the status
0 0 0 0
. They are LFT-verified and can be
used for billing.
Starting the system
Measuring test objects