ADJUSTMENT
6
6.4.9
One-point adjustment using the keypad (function CAL1)
When adjustment conditions (gas pressure, humidity and CO
2
concentration) are different
from the operating conditions of the measuring device, set the ambient parameters of the
measuring device to the adjustment environment for the adjustment duration. These
settings must then be changed back to represent the process conditions when the
TRANSIC111LP is reinstalled in its operating environment. Further information on setting
the ambient parameters of the TRANSIC111LP,
see “Setting ambient parameters”,
1
Check that no error messages are active.
Active error messages influence adjustment. Malfunction messages,
. (Serial interface) and
see “Display of active and undeleted errors (ERR)”,
(keypad) Error Table
2
Enter the password in menu
PAS
,
see “Entering the password (PAS)”, page 37
.
3
Select menu item
Cal1
. This holds the analog output.
4
Connect the span gas.
5
Enter the known O
2
value and confirm with
Ent
.
6
The measured value display blinks.
7
Feed adjustment gas.
8
Wait until the display shows a stable value.
9
Confirm with
Ent
.
PASS is displayed when the calibration is successful. The device now calculates the new
gain and offset settings and starts to display the new measured value.
10
Press
Ent
twice. This ends the one-point adjustment.
Fig. 38: One-point adjustment using the keypad
6.4.10
Restoring factory calibration
Restore the TRANSIC111LP factory calibration via the serial interface,
calibration (command FCRESTORE)”, page 59
, via the keypad,
.
Back
Back
k
<
<
<
<
<
<
<
!
Ent
Ent
2 x
<
<
Ent
Ent
Dn
Dn
Up
Up
Ent
Ent
Ent
Ent
<
5 s
The adjustment can be aborted at any time with
Back
.
The reference concentration used determines whether the gain or offset parameter
value is changed.
– Change to the offset value: Oxygen concentration < 10.5% O
2
– Change to the gain value: Oxygen concentration > 10.5% O
2