38
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
4
OPERATION
4.5.3
C0
2
content in process gas: Settings (CO2)
1
Enter the average CO
2
value in the sample gas.
Adjustable range: 0 ... 100% vol.
Fig. 25: CO
2
sample gas setting
4.5.4
One-point calibration (CAL1)
Figure can be found in Section Adjustment,
see “One-point adjustment using the keypad”,
.
4.5.5
Two-point calibration (CAL2)
Figure can be found in Section Adjustment,
see “Two-point adjustment using the keypad”,
.
4.5.6
Analog output display and settings (AOU)
1
Press
Ent
to display the current output value on the analog output.
2
To set fixed output values for the analog output (0, 4, 12, 20 mA), press Ent and use
Up
and
Dn
to select the analog output value.
Fig. 26: Displaying and setting the analog output value
.
4.5.7
Resetting to the factory calibration (FAC)
The adjustment is reset to factory setting.
(Gain value: 1, offset value: 0).
Fig. 27: Resetting to the factory setting for oxygen measurement
<
(0 ... 100 vol%)
<
<
<
<
<