93
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
TROUBLESHOOTING
8
36
NONFATAL
(non-serious)
ANALOG OUTPUT LOAD TOO
HIGH
Analog output load too high.
Check specifications for voltage supply unit and
cables. (See Technical Data,
37
NONFATAL
(non-serious)
NO MEASUREMENT RESULTS
No measuring results (results from other errors)
38
NONFATAL
(non-serious)
ANALOG OUTPUT RANGE
Oxygen concentration value measured outside set
output range.
Adjust the output range settings when necessary.
51
WARNING
SIGNAL QUITE LOW
Transmission (SIL) <20%
Maintenance request for optical components,
“Cleaning the optical components”, page 85
.
52
WARNING
EEPROM LOG&STATS COR-
RUPTED
Non-critical hardware error: EEPROM protocol and
statistics erroneous.
(Contact SICK Customer Service)
53
WARNING
WATCHDOG RESET OCCURRED Reset through software error.
Error
No.
Error category Error text
Cause
Table 13: Error Table