8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
4
OPERATION
4.11
Scaling and setting the analog output
4.11.1
Display/set output parameters (command OUT_PARAMS)
Syntax:
OUT_PARAMS
<cr>
Example:
4.11.2
Display/set pressure for compensation (command PRES)
1
Sets the pressure for compensation.
2
Use the SAVE command to store the setting in EEPROM,
.
Syntax:
PRES
[Pressure]<cr>
Set pressure for compensation (command XPRES)
This command is suitable for systems where the pressure value is measured continuously
and sent to the oxygen measuring device.
Syntax:
XPRES
[Pressure]<cr>
Example:
>out_params
NONFATALI (mA) : 3.000 ?
FATALI (mA) : 3.000 ?
I4 : 1 ?
OUTMAXO2 (%) : 20.000 ?
OUTMINO2 (%) : 0.000 ?
NONFATALI
=
Current output (in mA) for non-fatal error
FATALI
=
Current output (in mA) for fatal error
I4
=
Parameter to determine whether the current output range starts with 0 or 4 mA:
When I4 = 0, the current output is 0...20 mA
When I4 = 1, the current output is 4...20 mA
OUTMAXO2 (%)
=
Oxygen concentration OUTMAXO2 (%) is set for current output 20 mA
OUTMINO2 (%)
=
Oxygen concentration OUTMINO2 (%) is set for current output 0/4 mA
Pres-
sure
= Pressure of the measured gas (bar(a))
Saving the settings with command
SAVE
prevents the settings being lost at the next
reset.
>pres 1.300
PRESSURE(bar) : 1.300 ?
>save
EEPROM (basic) saved successfully
EEPROM (op) saved successfully
EEPROM (op_log1) saved successfully
EEPROM (op_log2) saved successfully
>
NOTICE:
This setting
CANNOT
be saved in EEPROM with command
SAVE
.
Pressure = Pressure setting (bar(a))
>xpres 1.300
PRESSURE(bar) : 1.300 ?