O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
INSTALLATION
3
3.1.3
Strong light sources near the oxygen measuring probe
The interfering effect is influenced by:
●
Filter used
●
Light incidence angle into the measuring probe
▸
Use a stainless steel filter for ambient light, e.g., normal interior or laboratory light.
▸
Use a PTFE filter for especially strong light sources, e.g. outdoors with direct sunlight.
3.1.4
Pressure
Observe the specifications on pressure conditions,
see “Flange mounted TRANSIC111LP-
A/-D/-F/-G/-H/-I/-J/-K”, page 14
see “TRANSIC111LP-C/-E with sample gas cell and wall
,
see “TRANSIC111LP-B for ambient gas measurement and wall
and
see “Dimensions and mechanics”, page 102
NOTICE:
Strong light sources interfere with TRANSIC111LP operation
▸
Prevent strong light sources reaching the measuring probe.