PRODUCT DESCRIPTION
2
2.3.3
Variant for ambient gas measurement
Fig. 4: TRANSIC111LP-B for ambient gas measurement and wall bracket
NOTICE:
The O
2
concentration in the transmitter is part of the O
2
measurement
When configuring ambient gas measurements, the complete TRANSIC111LP must be
exposed to the O
2
concentration to be measured.
The TRANSIC111LP variant for ambient measurements measures oxygen concentra-
tions from 2 ... 21 vol% O
2
. The measuring function is lost for oxygen concentrations
under 2 vol%.
Measuring environment
Measuring device complete in differing O
2
concentrations
0
2
: 2 … 25% 0
2
T:
−
40 … +80 °C
p: 0.8 … 1.4 bar(a)
SN 11058999
SUPPLY 11..36 V
Pmax 6VA
PN 1999998
TRANSIC111LP-1B1B1B1B1B
CLASS 1 LASER PRODUCT