82
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
6
ADJUSTMENT
The calibration starts. The following commands are now available for selection:
–
Enter
- output the most current measuring results or terminate continuous Print mode
–
R + Enter
- continuous output of measuring results with an interval of about 1 second.
Press
Enter
to terminate Print mode.
–
Esc
– to terminate the calibration
5
Wait until the measured value has stabilized. Enter the span gas concentration and
press
Enter
.
The new gain or offset parameter value is calculated and displayed.
The following is displayed after entering the command
COXY1
:
6
Enter command
SAVE
and press
Enter
. The new values are stored in EEPROM.
7
Enter command
ADJUST OFF
and press
Enter
.
>coxy1
Customer calibration
Current condition/settings:
Pressure (bar) : 1.013
H2O (g/m3) : 0
CO2 (vol-%) : 0
Gas temperature (C) : 23.64
Internal temperature (C): 24.84
If parameters are not correct, cancel calibration with ESC and change parameters
Connect ref gas to cuvette.
O2 (%): 20.52 Ref ?
O2 (%): 20.51 Ref ?
O2 (%): 20.51 Ref ? 20.50
Calibration data:
Pressure setting (bar) : 1.013
Measured oxygen : 20.51
Given oxygen : 20.50
Gas temperature (C) : 23.65
Ref path temperature (C) : 24.85
New Gain : 1.000
Calibration ready - remember SAVE command
>save
>save
EEPROM (op) saved successfully
EEPROM (op_log1) saved successfully
EEPROM (op_log2) saved successfully
>adjust off
Outputs to normal state
The adjustment is finished and the output returns to displaying the measurement
results.