![SICK TRANSIC111LP Operating Instructions Manual Download Page 102](http://html1.mh-extra.com/html/sick/transic111lp/transic111lp_operating-instructions-manual_1264716102.webp)
102
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
10
SPECIFICATIONS
10.2.4
TRANSIC111LP inputs and outputs
10.2.5
Dimensions and mechanics
Power supply (nominal)
Allowed input range
24 VDC, 500 mA
11 … 36 VDC
Power input
maximum
6 W
Maximum power consumption
U
in
= 11 VDC
U
in
= 24 VDC
550 mA
250 mA
Analog output
Maximum load
Precision
Temperature dependence
0/4 … 20 mA, source
500
Ω
±0.05% of full-scale value
±0.005%
⁄
°C
Serial output (2-wire, not isolated)
RS-485
Alarm/control relay
30 VAC , 1 A/ 60 VDC, 0.5 A
Serial output (NOTE: Only for maintenance)
RS-232
Connections
Screw terminals, 0.5...1.5 mm
2
RJ45 connection for RS-232
Resistance between signal ground and ground
10 M
Ω
Dimensions (H × W × D)
306 × 184 × 74 mm
3
Weight
2.2 kg
Enclosure material
G-AlSi10Mg (DIN 1725)
Enclosure classification
IP66
Flange
Can be attached to DIN/ANSI standard flanges.
Minimum flange sizes:
●
DIN EN 1092 DN50: Fitted with M16 DIN 933 or similar
●
ANSI ASME B16.5 (150) 2.5": Fitted with UNC 3
⁄
4"-10 or
similar
Cable bushing
Cable gland M20×1.5
Conduit screw fitting 1
⁄
2" NPT
Filter
Stainless steel mesh, holes 0.31 mm, wire thickness 0.2 mm
Hydrophobic PTFE filter, average pore size 8
μ
m
Materials with sample gas
contact
●
AISI 316 L(1.4404)
●
FKM or Kalrez
®
(optional)
●
PTFE, SiN, MgF2, quartz
●
Polymer coating