SETTING AMBIENT PARAMETERS
5
5
Setting ambient parameters
5.1
Compensation of ambient parameters
The TRANSIC111LP can compensate the temperature, pressure of the operating
environment as well as water and CO
2
content of the background gas.
The typical effect of the error depending on the process pressure is shown in the non-
compensated curve in
. The error magnitude is smallest at normal ambient air
pressure.
Fig. 31: Effect of process pressure compensation
0
Ambient parameters
Standard
Activated
Remarks
Operating pressure
(process pressure)
Standard ambient
parameters:
Pressure: 1 bar(a).
Must be activated,
ambient parameters
must be set.
The pressure outside the process in
which the measuring device enclosure
is installed should be normal ambient
air pressure. For further information,
see
Humidity
Water content
0 g
⁄
m
3
H
2
O
CO2
Relative carbon
dioxide concentra-
tion 0 vol% CO
2,
compensation is
deactivated.
Temperature
2 integrated
temperature sensors:
Inner temperature
Process temperature
Automatic, always
active
A significant difference between
process gas temperature and the
temperature in the measuring device
enclosure can influence the measured
value result.
Table 9: Compensation of ambient parameters
-
6
-
5
-
4
-
3
-
2
-
1
0
2
1
0.8
0.9
1.0
1.1
1.2
1.3
1.4
0
2
e
rror (% of measur
ed
value
)
Process pressure (bar(a))
x compensated at p = 1.2 bar
--- not compensated