Chapter
8
Operating Instructions
CLX490 Compact OMNI Scanner
8-12
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8009237/RB47/2007-08-03
Troubleshooting
1. The "Read Result" LED default setting:
Good Read) does not light up at the end
of the reading pulse:
– continued –
•
Evaluation criteria for bar code incorrect.
•
Bar code quality inadequate.
•
In CLV-Setup: choose the C
ODE
C
ONFIGU
-
RATION
tab, C
ODES
section. Enable all code
types (except Pharmacode), set code
lengths to F
REE
. Download temporarily to
CLX! Start P
ERCENTAGE
E
VALUATION
. Code
type and length are displayed in the
reading result. Enable recorded code
type in the C
ODES
section and disable all
other types. Enter the code length under
F
IXED
on the code type tab.
Download to CLX!
•
Check: sufficient blank zones present?
Print contrast adequate? Specified print
tolerances exceeded? In the "CLV-
Setup" program: choose R
EADING
C
ONFIURATION
tab, S
EGMENTATION
section.
Is S
TART
/S
TOP
selected? Perform read
with reference code.
2. The CLX is not transferring any reading
results to the host
•
Reading pulse is not functioning
correctly.
•
Wiring fault in data connection.
•
Voltage level incorrect.
•
Host interface in CLX parameterized
incorrectly.
•
See
Chapter 8.5.2 Malfunctions in Rea-
ding mode: reading trigger errors,
Page 8-10
•
Check wiring as shown in
•
Measure voltages.
•
In "CLV-Setup" program: choose H
OST
I
NTERFACE
tab, D
ATA
F
ORMAT
section. Are
the interface type (hardware) and data
format correct? Check settings in I
NTER
-
FACE
P
ROTOCOL
section.
3. The CLX transfers the status ST=3 to the
host in the reading result (if ST is enabled
for the separator; this is disabled by de-
fault)
The CLX has diagnosed a device error in the
self-test
Switch the CLX off and on again. Does the
"Device Ready" LED light up? If not, contact
the SICK Service department.
4. The data content of the bar code output
by the CLX is incorrect or incomplete
•
The data format on the host interface is
parameterized incorrectly in the CLX
•
The CLX suppresses the last character in
the bar code
•
In the "CLV-Setup" program: choose the
H
OST
I
NTERFACE
tab, D
ATA
F
ORMAT
section.
Are the values correct?
•
In the "CLV-Setup" program: choose the
C
ODE
C
ONFIGURATION
tab. Click the E
DIT
button. Choose the tab for the relevant
code. Is the T
RANSMIT
C
HECK
D
IGIT
check-
box activated? Change if necessary.
Download to CLX!
Malfunction
Possible cause
Remedy
Table 8-7: Troubleshooting: result output errors in Reading mode (contd.)