8.4.7 Protecting the Analyzer from Process Upsets
The analyzer should be protected from extended exposure to high concentrations of oxygen
or hostile gases. Automatically controlled solenoid valves should be installed to switch the
analyzer over to an N
2
purge when the process reaches some identifiable condition.
Gas line maintenance operations must also be examined for their effect on the analyzer.
For example, in many pipeline process or normal gas applications the plumbing system is
cleaned with either a liquid solvent or detergent solution. Since either causes damage to
the sensor, switch the analyzer over to a N
2
bypass purge, or shut off sample flow and
power to the analyzer prior to initiating the potentially hazardous process.