82 DF-760E
User Interface
Figure 72: Oxygen AutoSpan Recal
7.6.2.5 AutoSpan SetUp
Figure 73: Oxygen AutoSpan SetUp Screen
The AutoSpanSetUp screen allows for automatic span calibration or span checks of the
oxygen sensor at preset times. An optional warning period allows the user to disable the
AutoSpan function at critical measurement periods.
The
AutoSpan Interval
can be set for the frequency of spanning the oxygen sensor. The
date and time of the next scheduled AutoSpan is automatically calculated based on the
current AutoSpan Interval setting.
AutoSpan Timeout/Warning
: The timeout allows a time for stabilization. If the required
stable period is not achieved, the AutoSpan is automatically aborted. The Pre-Span
warning is set to provide an advance warning of a AutoSpan, based the minutes entered.
AutoSpan Type
: Allows the user to re-calibrate the oxygen sensor with each span or
merely to check if the span is within user specified limits.
A relay is available on the Oxygen Output Setup Screen (see page 97) to signal that a span
or zero calibration is taking place, and the analog output signal can also be frozen or
allowed to update during the calibration process.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...