60 DF-760E
User Interface
Figure 38: Moisture Zero Log Screen
7.5.5.2 System Error/Event Codes
The System Error Log reports functional errors and events related to the system. If the
error persists for more than 30 minutes, the code is displayed, if warranted. In addition, a
system alarm will trip if configured to do so. See page 64 for additional information on
setting System Alarms.
NOTE: The 30 minute clock is delayed for 60 minutes on a “cold” system start up.
Figure 39: System Error Log Screen
Following is a list of System Error/Event Codes and their descriptions:
101 = peak unstable or not found
133 = data acquisition system event
141 = sample gas pressure outside of pressure matrix range.
191 = fan condition voltage out of range – “Fan Failure”
Contact Delta F for assistance in interpreting the various codes if one should appear on the
screen.