User Interface DF-760E
97
7.6.7.5 Oxygen Sensor Off Alarm Setup
In the event that the oxygen sensor polarization voltage has been manually or
automatically turned off, an alarm can be enabled to indicate the current sensor status. For
additional information on control of the oxygen sensor see page 73. The user may assign
the sensor off alarm to one of four relays.
Figure 99: Oxygen Sensor-Off Alarm SetUp Screen
Use the
Next
key to move from between fields and use the arrow keys (
and
to
change the highlighted selections and to enter numerical values. When done, use the
Next
key to move to the
Accept
button and hit the
Enter
key to return to the main display.
Using the
ESC
key at anytime will exit the screen making no changes and return to the
main display.
7.6.8 Oxygen Analog Output Setup
Figure 100: Oxygen Analog Output Setup Menu
The
Zero Point
corresponds to the lowest voltage or current output (0 VDC, 4 mA) that is
sent to a recorder, while the
Full Scale
corresponds to the maximum voltage or current
output (1/5/10 VDC or 20mA) that is sent. The Full Scale set point (FS) is set from 0.002
ppm to 20.00 ppm.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...