Connecting to External Devices DF-760E
41
6 Connecting to External Devices
The analyzer can be interfaced to a variety of external devices via the ports on the rear
panel. Alarm contacts, voltage, and current outputs, and serial communications are
supported. All outputs, analog or digital, are fully isolated from earth ground.
NOTE
During the six minute warm up period at startup, all analog and
digital outputs are held to an artificial 0.011ppb reading to avoid the
reporting of false readings.
Figure 13: Rear Panel Electrical Connectors
6.1
Serial Communication Port – J5
Either of two serial communication ports are available at the time of order: RS232C or
RS485 which enable interface between the analyzer and other operating systems.
Up to 32 units may be accessed via RS-485. Operating parameters are 8 bits, no parity,
and one stop bit. Baud rate may be selected from the menu on the display.
See Table 2 and Table 3 for wiring information.
J8
J9
J2
J3
J10
J5
J4
OXYGEN ANALOG
OUTPUT
SERIAL COMMUNICATION
MOISTURE ANALOG OUTPUT
RELAY OUTPUTS
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...