User Interface DF-760E
111
used to choose the appropriate relay number. The
NEXT
key is then used to move to the
Activate/Deactivate
field where the arrow keys (
and
) are used to toggle between
the two options. The
NEXT
key is then used to move to the
Apply
field where the
Enter
key is hit to change the relay state. This process can be repeated as often as needed. When
done, the
NEXT
key is used to move to the
Done
field and the
Enter
key is hit to leave
the screen. The condition of the relays before the test is restored when the test is
concluded. See the section on relay ports found on page 43 for additional information.
Figure 127: Test Oxygen Relay
7.7.8 Test Analog Outputs
Figure 128: Test Analog Outputs Menu
7.7.8.1 Test Moisture Analog Output
The Test Output screen, as shown in Figure 129 is used to calibrate the moisture analog
recorder output. When the Test Output screen is selected, the
NEXT
key is used to move
to the percentage field where the arrow keys (
and
) are used to choose the
appropriate setting. The
NEXT
key is then used to move to the
Apply
field where the
Enter
key is hit to set the analog output to the selected value. The analog output response
should match the value that was entered. For example, if 80 percent is entered for the
percent full scale level, and the analog output is set for 0 to 10 VDC, the analog output is
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...