86 DF-760E
User Interface
7.6.3 Maintenance
Figure 78: Oxygen Maintenance Menu
7.6.3.1 Log a Replenishment Solution Addition
This command, allows the user to acknowledge that replenishment solution has been added
to the oxygen sensor and if the “Replenish Sol’n” reminder flag is active on the display, it
will be cleared. Solution additions are tracked in the Replenish Sol’n Log as shown on
page 92. The current date is then used to automatically calculate the date for the next
solution addition reminder based on the criteria as set on the Replenish Sol’n Reminder
Setup screen below.
7.6.3.2
Replenish Sol’n Reminder Setup
The Add Replenishment Solution reminder can be turned on or off and a time period (in
weeks) can be set for solution addition reminders.
If the electrolyte level is low in the oxygen sensor, only Delta F
Replenishment Solution should be added to the sensor.
Do not add
electrolyte solution to restore the electrolyte level.
Do not overfill.
Figure 79: Replenish Sol’n Reminder Setup Screen
7.6.3.3
Clear Replenish Sol’n Reminder
The Replenish Sol’n Reminder flag can be cleared from the main data display without
indicating in the Replenish Sol’n Addition log that solution has been added to the oxygen
sensor.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...