User Interface DF-760E
63
Figure 42: Moisture Alarm Setup Screen
7.5.6.2 Moisture Temperature Range Alarm Setup
Figure 43: Moisture Temperature Alarm Setup
The system is constantly monitoring the ambient temperature in the analyzer cabinet. If
enabled on the Temperature Alarm Setup screen, an alarm can be assigned to trip if the
ambient temperature exceeds preset limits. The user may assign the temperature alarm to
one of four relays.
Use the
Next
key to move from between fields and use the arrow keys (
and
to
change the highlighted selections and to enter numerical values. When done, use the
Next
key to move to the
Accept
button and hit the
Enter
key to return to the main display.
Using the
ESC
key at anytime will exit the screen making no changes and return to the
main display.
7.5.6.3 Moisture Pressure Range Alarm Setup
The system is constantly monitoring the pressure in the gas path and the result is displayed
on the front panel. If enabled on the Pressure Alarm Setup screen, an alarm can be assigned