Installation, Start Up and Shut Down DF-760E
27
CAUTION
A 220 VAC power cord is not supplied with the analyzer. The customer supplied
220 VAC power cord should be rated for 250 VAC, 10 Amps and be equipped
with an IEC 60320 C13 connector for connecting to the analyzer. This power
cord provides the protective ground for the analyzer so it must have a protective
earth wire. Ensure that the power cord is plugged into a properly grounded
outlet. The power cord should have the appropriate safety agency approvals for
your location.
Plug the line cord (supplied with 110 VAC units only) into the receptacle at the back of the
analyzer. Connect the line cord to the power source.
4.1.9 Hydrogen Service Safety System
This optional system is designed to safeguard the DF-760 from explosion hazards when
operating on hydrogen sample gas under normal pressure and flow conditions as detailed
in the Operating Instruction Manual. The instrument chassis and the remote pump, if
equipped, are both protected by maintaining a safe condition within their respective
enclosures. If installed, this option impacts the electrical wiring, gas plumbing and
operation of the analyzer. See page 173 for additional installation and operation
information.
NOTE, if equipped with the Hydrogen Safety Service System, when shipped from the
factory the analyzer will be configured through the GSF screen to measure hydrogen. As a
result, the Hydrogen Safety Service System will be enabled out of the box.
Figure 5: AC Power Input
GND
RLY5-NC
RLY6-COM
RLY6-NO
RLY6-NC
RLY5-COM
RLY5-NO
KEY
5
5
RLY4-COM
RLY2-COM
RLY1-COM
RLY1-NO
RLY1-NC
KEY
4
4
RLY3-NC
1
2
3
KEY
3
2
1
RLY3-COM
RLY3-NO
RLY2-NO
RLY2-NC
GND
8
6
7
GND
8
7
6
RLY4-NO
RLY4-NC
5
5
4
KEY
4
1
2
3
1
2
3
RLY7-COM
8
GND
8
6
7
6
7
RLY8-NC
RLY8-NO
RLY8-COM
RLY7-NC
RLY7-NO
RX+
TX+
3
2
1
4-20mA-
4-20mA+
3
3
1
2
1
2
4-20mA-
4-20mA+
TX-
GND
RX-
KEY
7
8
5
4
6
AOUT+
GND
KEY
AOUT-
8
7
6
4
5
KEY
GND
8
7
6
4
5
AOUT-
AOUT+
~
~
~
~
~
~
AC INLET
VACUUM PUMP AC POWER OUT
VACUUM PUMP ON/OFF
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...